Compatibility Language - F
181
Figure F-1. COMpatibility Status Model
Table F-4. Bit Assignment of Status, Astatus, Fault, & Mask Registers
Bit Position 11 10 9876543210
Bit Name NORM FAST -CC INH ERR OC OS OT OV UNR +CC CV
Bit Weight 2048 1024 512 256 128 64 32 16 8421
CV = The unit is operating in constant voltage mode.
+CC = The unit is operating in constant current mode.
UNR = The output of the unit is unregulated.
OV = The overvoltage protection circuit has tripped.
OT = The over-temperature protection circuit has tripped.
OC = The overcurrent protection circuit has tripped.
OS = Open sense lead detected (Agilent 66111A, 66311B/D)
ERR = A programming error has occurred. Use ERR? to clear.
-CC = The unit is operating in negative constant current mode.
INH = The external remote inhibit signal has turned the output off.
FAST = The output is in Fast operating mode. (Agilent 6632B, 6633B, 6634B only)
NORM = The output is in Normal operating mode. (Agilent 6632B, 6633B, 6634B only)
Table F-5. Bit Configuration of Serial Poll Register
Bit Position 76543210
Bit Name not used RQS ERR RDY no t used not used PON FAU
Bit Weight 64 32 16 2 1
RQS = The dc source has generated a service request. Use a serial poll to clear.
ERR = Same as ERR bit in Status register. Use ERR? to clear.
RDY = This bit cleared when unit busy processing commands. Set when processing complete.
PON = A Power-on has occurred. Use CLR to clear.
FAU = A bit has been set in the Fault register. Use FAULT? to clear.
STATUS
1
2
16
64
128
CV
+CC
UNR
OT
OS
OC
ERR
INH
OV
4
512
-CC
FAST
8
32
256
1024
2048
REGISTER
NORM
STATUS
1
2
16
64
128
4
512
8
32
256
1024
2048
REGISTER
ACCUMULATED MASK
1
2
16
64
128
4
512
8
32
256
1024
2048
REGISTER
CV
+CC
UNR
OT
OS
OC
ERR
INH
OV
-CC
FAST
NORM
FAULT
1
2
16
64
128
4
512
8
32
256
1024
2048
REGISTER
SERIAL
1
2
16
64
128
4
8
32
REGISTER
POLL
not used
not used
not used
PON
FAU
RDY
ERR
RQS