Verification and Calibration - B
153
Table B-4. Verification Test Record for Agilent Model 66309B/D
Model Agilent___________ Report No.____________Date_____________
Test Description Minimum
Specification Recorded Results M aximum
Specification
Voltage Programming and Measurement
Low Voltage Vlo 10 mV _____mV +10 mV
Front Panel measurement Vfplo Vlo 5 mV _____mV Vlo +5 mV
High Voltage Vhi 14.982 V _______V 15.018 V
Front Panel measurement Vfphi Vhi 9.5 mV _______V Vhi +9.5 mV
Outp 2 Low Voltage V2lo 40 mV _____mV +40 mV
Outp 2 Front Panel measurement V2fplo V2lo 15 mV _____mV V2lo +15 mV
Outp 2 High Voltage V2hi 11.936 V _______V 12.064 V
Outp 2 Front Panel measurement V2fphi V2hi 39 mV _______V V2hi +39 mV
Current Programming and Measurement (High Range)
Low Current Ilo 1.33 mA _____mA 1.33 mA
High Current Ihi 2.9972 A _______A 3.0028 A
Front Panel measurement Ifphi Ihi 6.5 mA _______A Ihi +6.5 mA
Outp 2 Low Current I2lo 4.5 mA _____mA 4.5 mA
Outp 2 High Current I2hi 1.492 A _______A 1.508 A
Outp 2 Front Panel measurement I2fphi I2hi 6 mA _______A I2hi +6 mA
Current Measurement (Low Range)
Low Current measurement Ilrlo Ilrlo 2.5 µA _____µA Ilrlo +2.5 µA
High Current measurement Ilrhi Ilrhi 22.5 µA_____mA Ilrhi +22.5 µA
Current Sink Measurement
Low Current Sink measurement Isinklo Isinklo 22.5 µA_____mA Isinklo +22.5 µA
High Current Sink measurement I sinkhi Isinkhi 5.1 mA _____A Isinkhi +5.1 mA
DVM Voltage Measurement (Agilent 66309D only)
Positive Voltage measurement Dvpos Dvpos 15 mV _____V Dvpos +15 mV
Negative Voltage measurement Dvneg Dvneg 6.8 mV _____V Dvneg +6.8 mV