A - Specifications

Table A-2. Supplemental Characteristics (continued)

Parameter

Recommended Calibration Interval

Regulatory Compliance

 

All Models

from the date the unit is put into

1 year

service

 

Listing pending:

UL 3111-1

Certified to:

CSA 22.2 No. 1010.1

Conforms to:

IEC 1010-1, EN 61010-1

Complies with:

EMC directive 89/336/EEC (ISM Group1

 

Class B)

Dimensions (see figure 3-1)

Net weight

Shipping weight

Height:

Width:

Depth:

88.1mm (3.5in.) 212.8 mm (8.4in.)

435 mm (17.125 in.)

9.07kg (20 lbs.)

11.1kg (24.5 lbs.)

Table A-3. Agilent 66311D/66309D DVM Specifications and Characteristics

Performance Specifications (warranted from 0 to 55 °C unless otherwise noted)

DC Readback Accuracy (@ 25°C ±5°C)

0.04% +5 mV

AC+DC rms Readback Accuracy

60 Hz to 10 kHz:

(@ 25°C ±5°C with dc plus a sinewave input > 25mV rms)

45-60 Hz and 10-20 kHz:

Common Mode Voltage Range (from either DVM input with

 

respect to the negative output terminal of Output 1)

 

Maximum DC Differential Voltage

 

Maximum AC Differential Voltage (with a sinewave input)

 

Supplemental Characteristics

1% +5 mV1

4% +5 mV1

4.5 Vpk to + 25 Vpk

±25 V peak

10 V rms2

Maximum Continuous Input Capability without damage (between input terminals or from either input to chassis ground)

50 V

Input Resistance (from either DVM input with respect to either output terminal of Output 1)

Input Capacitance (on either input terminal)

DC Common Mode Rejection Ratio

Voltage Readback Resolution

Readback Temperature Coefficient (change per °C)

Front panel: GPIB:

20 MΩ

<60 pF

>83 dB

1mV

<0.2 mV

0.002% + 0.2 mV

1+15 mV for dc plus sinewave input < 25 mV rms.

2To accept 10 Vrms sinewave input, the common mode voltage with respect to the negative terminal of output 1 must be

10Vdc. This is required to "center" the DVM in its common mode range.

Table A-4. Agilent 66309B/D Option 521 Characteristics

Output Impedance

Output 1:

(Output = OFF)

Output 2:

Solid State Relay Current rise time

Relay mode =Hot

(from 10% to 90% of the total output change)

 

Solid State Relay Current fall time

Relay mode =Hot

(from 10% to 90% of the total output change)

 

500k ohms

200k ohms

100microseconds

50 microseconds

146

Page 146
Image 146
Agilent Technologies 66311B, 66111A manual 146, Table A-4. Agilent 66309B/D Option 521 Characteristics

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.