Front Panel Operation - 5
53
Metering Keys
Metering keys control the metering functions of the dc source. As set from the factory, all front panel
measurements from the main output (output 1), are calculated from a total of 2048 readings taken at a
46.8 microsecond sampling rate. Therefore, the factory default acquisition time for a single front panel
measurement is about 100 milliseconds. Refer to “Making Front Panel Measurements” for more
information about changing the front panel sampling rate and the number of measurement points.
All front panel measurements from the DVM and from output2 are fixed at 2048 measurement readings
taken at a 15.6 microsecond sampling rate.
NOTE: The front panel sample rate and data point settings are separate and independent of the
sample rate and data point settings that are programmed over the GPIB interface. When
an GPIB measurement is in progress, the front panel display temporarily indicates
-- -- -- -- --. Front panel measurements resume when the GPIB measurement completes.
Press this key to access the meter menu list. Also use this key to exit a
menu at any time and return to meter mode.
Display Measurement
<reading>V <reading>A Measures output dc voltage and current
<reading>V MAX Measures peak output voltage1
<reading>V MIN Measures minimum output voltage1
<reading>V HIGH Measures the high level of a volta ge waveform1
<reading>V LOW Measures the low level of a voltage waveform1
<reading>V RMS Measures rms voltage1
<reading>A MAX Measures peak output current1
<reading>A MIN Measures minimum output current1
<reading>A HIGH Measures the high level of a current waveform1
<reading>A LOW Measures the low level of a current waveform1
<reading>A RMS Measures rms current1
<reading>V DC:DVM Measures dc voltage on DVM input2
<reading>V RMS:DVM Measures rms voltage on DVM input2
Press this key to access the following metering functions.
Display Command Function
CURR:RANGE <char> Select current range (AUTO | LOW | HIGH1)
CURR:DET <char> Select current measurement bandwidth (ACDC | DC1)
TINT <value> Sets the front panel measurement interval in seconds
(15.6 µs to 1 second)
POINTS <char> Sets the # of points in front panel measurement buffer
( 1, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048)
Notes: 1not valid for Agilent 66111A
2only valid for Agilent 66311D/66309D
reading = the returned measurement
value = a numeric value
char = a character string parameter
Use and to scroll through the menu commands.
Use and to scroll through the menu parameters.
Use and to select a digit in a numeric entry field.
Meter
Shift Input