4

Turn-On Checkout

Checkout Procedure

Successful tests in this chapter provide a high degree of confidence that your unit is operating properly. For verification tests, see appendix B. Complete performance tests are given in the Service Guide.

NOTE: To perform the checkout procedure, you will need a wire for shorting the output terminals together.

The following procedure assumes that the unit turns on in the factory-default state. If you need more information about the factory default state, refer to the *RST command in chapter 8. Note that the values shown in the Display column may not exactly match the values that appear on the front panel of the unit.

If you have not already done so, connect the power cord to the unit and plug it in. Connect the output connector to the back of the unit with the sense jumpers installed.

 

Procedure

Display

Explanation

1.

Turn the unit on. The dc

**********

During selftest, all display segments are briefly lit,

 

source undergoes a self-test

ADDRESS 5

followed by the GPIB Address.

 

when you first turn it on.

0.000V 0.0001A

The display then goes into meter mode with the Dis

 

 

annunciator on, and all others off. In Meter mode the n.nnnV digits indicate the output voltage and the

.nnnnA digits indicate the output current. The flashing digit on the display indicates the digit that will be affected if changes are made to the displayed values using the rotary control or the and keys. You will only see the changes if the output is ON.

NOTE: Press the Meter key to exit a menu at any time and return to meter mode. If the Err annunciator on the display is on, press the Shift key followed by the Error key to see the error number. See table 4-1 at the end of this chapter.

2.Check that the fan is on.

You should be able to hear the fan and feel the air coming from the back of the unit.

3.

Unplug the output connector

-0.224V

0.0000A

The output voltage indicates approximately -0.2 volts

 

from the back of the unit.

 

 

because the output sense connections have opened.

4

Press Output, scroll to

SENSE:PROT ON

Enables the open sense detect circuit.

 

SENSE:PROT and select

 

 

 

 

ON. Press Enter

 

 

 

5.

Press Output On/Off

-0.224V

0.0000A

The open sense detect circuit disables the output. The

 

 

 

 

Dis annunciator is off, but the Prot annunciator is on.

6.

Press Protect

+/- SENSE OPEN

Display indicates the protection condition.

45

Page 45
Image 45
Agilent Technologies 66311B, 66309B Checkout Procedure, Procedure Display Explanation, Press Output On/Off, Press Protect

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.