8 – Language Dictionar y
128
TRIGger:SEQuence2TRIGger:ACQuire
These commands generate a BUS trigger for the measurement trigger system. When the measurement
trigger system is enabled, the measurement trigger causes the dc source to measure either the voltage or
current on the main output or the DVM inputs and store the results in a buffer. The SENSe:FUNCtion
command selects the signal that will be measured.
Command Syntax TRIGger:SEQuence2[:IMMediate]
TRIGger:ACQuire:[:IMMediate]
Parameters None
Examples TRIG:SEQ2 TRIG:ACQ
Related Commands TRIG:SOUR TRIG:SEQ2:DEF TRIG:SEQ2:COUN
TRIG:SEQ2:LEV:VOLT TRIG:SEQ2:SLOP:CURR
SENS:FUNC
TRIGger:SEQuence2:COUNt:CURRentTRIGger:ACQuire:COUNt:CURRent
This command sets up a successive number of triggers for measuring current data. With this command,
the trigger system needs to be initialized only once at the start of the acquisition period. After each
completed measurement, the instrument waits for the next valid trigger condition to start another
measurement. This continues until the count has completed.
Command Syntax TRIGger:SEQuence2:COUNt:CURRent<NRf+>
TRIGger:ACQuire:COUNt:CURRent<NRf+>
Parameters 1 to 100
*RST Value 1
Examples TRIG:SEQ2:COUN:CURR 5 TRIG:ACQ:COUN:CURR 1
Query Syntax TRIGger:SEQuence2:COUNt:CURRent?
TRIGger:ACQuire:COUNt:CURRent?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2 TRIG:ACQ
TRIGger:SEQuence2:COUNt:DVMTRIGger:ACQuire:COUNt:DVM
Agilent 66311D/66309D only
This command specifies a DVM trigger count. Since the trigger count for DVM measurements cannot be
set to any value other than 1, this command does not need to be used. It is included for completeness
only.
Command Syntax TRIGger:SEQuence2:COUNt:DVM<NRf+>
TRIGger:ACQuire:COUNt:DVM<NRf+>
Parameters 1
*RST Value 1
Examples TRIG:ACQ:COUN:DVM 1
Query Syntax TRIGger:SEQuence2:COUNt:DVM?
TRIGger:ACQuire:COUNt:DVM?
Returned Parameters <NR3>