A - Specifications

Supplemental Characteristics

Table A-2 lists the supplemental characteristics, which are not warranted but are descriptions of typical performance determined either by design or type testing.

Parameter

Output Programming Range

Average Programming Resolution

OVP Accuracy

Table A-2. Supplemental Characteristics

 

Agilent 66111A

 

Agilent

 

 

66311B/D,

 

 

Agilent 66309B/D

 

 

output 1 only

Voltage:

0 – 15.535 V

 

Current:

0 – 3.0712 A

 

OVP:

22 V

 

Voltage:

3.75 mV

 

Current:

0.75 mA

 

OVP:

100 mV

 

 

2.4 % + 240 mV

 

 

 

 

Agilent

66309B/D

output 2 only

0 – 12.25 V

0 – 1.52 A

NA

14mV 1.75 mA

NA

NA

Average Current Measurement Resolution

Maximum Current Measurement

Sink Current

Programming Accuracy Temperature Coefficient (change/C°)

Readback Accuracy Temperature Coefficient (change/C°)

Drift2

Remote Sense Capability

(voltage per load lead)

High Range:

Low Range:

Voltage:

Current:

OVP:

Voltage: Current (dc):

Current (acdc): Current (low range):

Voltage:

Current: +

3.5 mA

213 μA

NA

0.6 μA

7A

-2 A @ 7.5 V1

0.01% + 0.5 mV

0.01% + 15 μA

0.015% + 4 mV

0.01% + 300 μV

0.01% + 300

μV

0.02% + 15 μA

0.02%

+ 15 μA

NA

0.05% + 120 μA

NA

0.01%

+ 0.3

μA

0.01% + 1 mV

0.01% + 30 μA

up to 4 V3

55μA

NA

1.8A

- 0.03 A @ 7.5 V

0.03% + 1.5 mV

0.03% + 0.4 mA

NA

0.02% + 1.5 mV

0.02% + 0.4 mA

NA

NA

0.03% + 3 mV

0.03% + 0.8 mA

up to 4 V4

Output Voltage Rise Time Output Voltage Fall Time

Output Voltage Settling Time5

Dynamic Measurement System 6

10% to 90%:

90% to 10%:

0 to rated voltage: rated voltage to 0:

Buffer Length: Sample Rate Range:

 

< 200 μs

 

< 200 μs

 

0.5 ms

 

1 ms

NA

 

0 4096 points

 

 

15.6μs 31200s

<200 μs

<35 ms

NA NA

NA

1The sink current decreases linearly from 2.8A @ 0 V to 1.2 A @ 15 V. Sink current does not track the programmed current.

2Following a 30 minute warm-up, the change in output over 8 hours, under constant ambient, load and line conditions.

3Add 2 mV to the load regulation for each 1 V change in the positive output lead.

4Add 3 mV to the load regulation for each 1 V change in the negative output lead.

5To settle within 1 LSB or 3.75mV of the final value for Output 1.

6For a pulse waveform, the accuracy of any individual data point in the buffer is dependent on the rise time of the pulse. For a current pulse of 1.4A with a rise time constant of 50μs, the error in measurement of a single data point during the pulse rise time is typically 10mA.

144

Page 144
Image 144
Agilent Technologies 66311B manual 144, Table A-2. Supplemental Characteristics, Agilent 66309B/D Output 2 only

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.