Specifications

Specifications

Table A-1 lists the specifications of the dc source. Unless otherwise noted, specifications are warranted
over the ambient temperature range of 0 to 55 °C. Specifications apply with typical cellular phone
capacitive loads from 0µF to 12,000µF. Sensing is at the rear terminals of the power supply after a 30-
minute warm-up period. Sense terminals are externally jumpered to their respective output terminals.
Table A-1. Performance Specifications
Parameter Agilent 66111A Agilent 66311B/D;
Agilent 66309B/D
output 1 only
Agilent
66309B/D
output 2 only
Output Ratings Voltage:
Current:
Peak Current:
0 – 15 V
0 – 3 A
5 A1
0 – 12 V
0 – 1.5 A
2.5 A2
Programming Accuracy
(@ 25°C ±5°C) Voltage:
+Current: 0.05% + 10 mV
0.05% + 1.33 mA30.2% + 40 mV
0.2% + 4.5
mA
DC Measurement Accuracy
(via GPIB or front panel meters with
respect to actual output @ 25°C
±5°C)
Voltage:
Current:
High Current range4
+20 mA to + rated I:
20 mA to rated I:
Low Current range
20 mA to +20 mA:
0.03% + 5 mV
see below
0.2% + 9 mA
0.2% + 9 mA
NA
0.03% + 5 mV
see below
0.2% + 0.5 mA5
0.2% + 1.1 mA
0.1% + 2.5 µA6
0.2% + 15 mV
0.2% + 3 mA
NA
NA
NA
Ripple and Noise
(in the range of 20 Hz to 20 MHz
with outputs ungrounded or with
either terminal grounded)
Voltage (rms/p-p):
Current (rms): 1 mV/ 6 mV7
2 mA 1 mV/6 mV7
2 mA
Load Regulation
(change in output voltage or current
for any load change within ratings)
Voltage:
Current: 2 mV
0.75 mA 1.6 mV
0.375 mA
Line Regulation
(change in output voltage or current
for any line change within ratings)
Voltage:
Current: 0.5 mV
0.75 mA 0.4 mV
0.25 mA
Transient Response Time
(for the output voltage to recover to
20 mV below its final setting)
< 35 µs8< 400 µs9
1 Peak current for up to a 7 millisecond time period. Average current cannot exceed 3 A.
2 Peak current for up to a 1 millisecond time period. Average current cannot exceed 1.5 A.
3 0.05% +2.3mA when programming between zero and 0.03% of full scale current.
4 Agilent 66111A units do not have a low current range, the dc measurement accuracy applies from 0 amperes to the rated current.
5 Applies with current detector set to DC. ACDC mode accuracy is 0.2% + 3mA for sourcing and 0.2% + 3.6 mA for sinking.
6 This specification may degrade slightly when the unit is subjected to an RF field 3 V/meter.
7 Specification is for phone capacitance greater than 6µF.
8 Applies to the negative portion of the output voltage transient waveform, following a 0.1 A to 1.5 A load current change.
(Applies with unit set to High capacitance compensation range with 5 µF minimum output load capacitance.)
9 Following a 0.75 A to 1.5 A load change.