Agilent Technologies 66311B, 66309B MEASureVOLTage2, MEASureVOLTageACDC? FETChVOLTageACDC?, 109

Models: 66111A 66309B 66311B D

1 190
Download 190 pages 33.49 Kb
Page 109
Image 109

Language Dictionary - 8

MEASure:VOLTage2

Agilent 66309B/D only

This query measures the output voltage at the auxiliary output. Output 2 measurements are calculated from a total of 2048 readings taken at a 15.6 microsecond sampling rate. These parameters are fixed.

Query Syntax

MEASure[:SCALar]:VOLTage2[:DC]?

Parameters

None

Examples

MEAS:VOLT2? FETC:VOLT2:DC?

Returned Parameters

<NR3>

Related Commands

MEAS:CURR2?

MEASure:VOLTage:ACDC?

FETCh:VOLTage:ACDC?

Agilent 66311B/D, 66309B/D only

These queries return the ac+dc rms output voltage.

Query Syntax

MEASure[:SCALar]:VOLTage:ACDC?

 

FETCh[:SCALar]:VOLTage:ACDC?

Parameters

None

 

Examples

MEAS:VOLT:ACDC?

FETC:VOLT:ACDC?

Returned Parameters

<NR3>

 

Related Commands

MEAS:CURR:ACDC?

 

MEASure:VOLTage:HIGH?

FETCh:VOLTage:HIGH?

Agilent 66311B/D, 66309B/D only

These queries return the High level voltage of a voltage pulse waveform. The instrument first measures the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse waveform using 16 bins between the maximum and minimum data points. The bin containing the most data points above the 50% point is the high bin. The average of all the data points in the high bin is returned as the High level. If no high bin contains more than 1.25% of the total number of acquired points, then the maximum value is returned by these queries.

Query Syntax

MEASure[:SCALar]:VOLTage:HIGH?

 

FETCh[:SCALar]:VOLTage:HIGH?

Parameters

None

 

Examples

MEAS:VOLT:HIGH?

FETC:VOLT:HIGH?

Returned Parameters

<NR3>

 

Related Commands

MEAS:VOLT:LOW?

CALC:REF:HIGH

109

Page 109
Image 109
Agilent Technologies 66311B MEASureVOLTage2, MEASureVOLTageACDC? FETChVOLTageACDC?, MEASureVOLTageHIGH? FETChVOLTageHIGH?