Language Dictionary - 8

Command Syntax OUTPut[12]:RELay:MODE <mode>

Parameters DD DH HD HH

Examples OUTP:REL:MODE DH (sets output 1 relay mode DH)

OUTP2:REL:MODE HH (sets output 2 relay mode HH)

Query Syntax OUTPut[12]:REL:MODE?

Returned Parameters <CRD>

CAUTION: Non-volatile memory has a finite maximum number of write cycles. Programs that repeatedly cause write cycles to non-volatile memory can eventually exceed the maximum number of write cycles and cause the memory to fail.

OUTPut:DFI

This command enables or disables the discrete fault indicator (DFI) output from the dc source.

Command Syntax

OUTPut:DFI[:STATe] <bool>

Parameters

0 1 OFF ON

 

*RST Value

OFF

 

Examples

OUTP:DFI 1

OUTP:DFI ON

Query Syntax

OUTPut:DFI[:STATe]?

Returned Parameters

0 1

 

Related Commands

OUTP:DFI:SOUR

 

OUTPut:DFI:SOURce

This command selects the source for discrete fault indicator (DFI) events. The choices are:

QUEStionable

selects the Questionable event summary bit (bit 3 of the Status Byte Register)

OPERation

selects the Operation Event summary bit (bit 7 of the Status Byte Register)

ESB

selects the Standard Event summary bit (bit 5 of the Status Byte Register)

RQS

selects the Request Service bit (bit 6 of the Status Byte Register)

OFF

selects no DFI source

Command Syntax

OUTP:DFI:SOUR <source>

Parameters

QUES OPER ESB RQS OFF

*RST Value

OFF

Examples

OUTP:DFI:SOUR OPER

Query Syntax

OUTPut:DFI:SOUR?

Returned Parameters

<CRD>

Related Commands

OUTP:DFI

OUTPut:PON:STATe

This command selects the power-on state of the dc source. This information is saved in non-volatile memory. The following states can be selected:

RST Sets the power-on state to *RST. Refer to *RST for more information.

RCL0 Sets the power-on state to *RCL 0. Refer to *RCL for more information.

Command Syntax

OUTPut:PON:STATe <state>

Parameters

RST RCL0

Examples

OUTP:PON:STAT RST

Query Syntax

OUTPut:PON:STATe?

Returned Parameters

<CRD>

Related Commands

*RST *RCL

115

Page 115
Image 115
Agilent Technologies 66111A, 66311B, 66309B manual OUTPutDFISOURce, OUTPutPONSTATe, 115

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.