Language Dictionar y - 8
133
TRIGger:SEQuence2:SLOPe:DVMTRIGger:ACQuire:SLOPe:DVM
Agilent 66311D/66309D only
This command sets the slope of the DVM input signal.
POSitive measurement triggering occurs on the rising edge.
NEGative measurement triggering occurs on the falling edge.
EITHer measurement triggering occurs on either edge.
Command Syntax TRIGger:SEQuence2:SLOPe:DVM<slope>
TRIGger:ACQuire:SLOPe:DVM<slope>
Parameters EITHer | POSitive | NEGative
*RST Value POSitive
Examples TRIG:SEQ2:SLOP:DVM POS TRIG:ACQ:SLOP:DVM EITH
Query Syntax TRIGger:SEQuence2:SLOPe:DVM?
TRIGger:ACQuire:SLOPe:DVM?
Returned Parameters <CRD>
Related Commands TRIG:SEQ2:LEV:DVM
TRIGger:SEQuence2:SLOPe:VOLTageTRIGger:ACQuire:SLOPe:VOLTage
This command sets the slope of an internally triggered voltage measurement.
POSitive triggering occurs on the rising edge.
NEGative triggering occurs on the falling edge.
EITHer triggering occurs on either edge.
Command Syntax TRIGger:SEQuence2:SLOPe:VOLTage<slope>
TRIGger:ACQuire:SLOPe:VOLTage<slope>
Parameters EITHer | POSitive | NEGative
*RST Value POSitive
Examples TRIG:SEQ2:SLOP:VOLT POS TRIG:ACQ:SLOP:VOLT EITH
Query Syntax TRIGger:SEQuence2:SLOPe:VOLTage?
TRIGger:ACQuire:SLOPe:VOLTage?
Returned Parameters <CRD>
Related Commands TRIG:SEQ2:SLOP:CURR