Language Dictionar y - 8
111
SENSe:CURRent:DETector
Agilent 66311B/D, 66309B/D only
This command lets you select the type of detector used for output current measurements. Two choices for
detecting current measurements are available:
ACDC This is the preferred choice for all dynamic current measurements. When ACDC is selected,
the measured output current includes the current that flows in the instrument’s output
capacitor. It is especially important to use ACDC detection when measuring pulse or other
waveforms with frequency contents greater than several kilohertz.
DC Select DC only if you are making dc current measurements and you require a dc measurement
offset accuracy better than 2mA on the High current measurement range. When DC is
selected, the component of output current that is supplied by the instrument’s output filter is
not sensed. Note that this selection gives inaccurate results on current waveforms with
frequency contents greater than several kilohertz.
NOTE: This command only applies to the High current measurement range.
Command Syntax SENSe:CURRent:DETector <detector>
Parameters ACDC or DC
*RST Value ACDC
Examples SENS:CURR:DET ACDC SENS:CURR:DET DC
Query Syntax SENSe:CURRent:DETector?
Returned Parameters <CRD>
SENSe:CURRent:RANGe
Agilent 66311B/D, 66309B/D only
This command selects the dc current measurement range:
High Range: 0 through MAX (see Table 8-3)
Low Range: 0 through 0.02 A
The High range covers the full current measurement capability of the instrument. The Low range
measures currents up to a maximum of 20 mA. This increases the low current measurement sensitivity
for greater accuracy and resolution. The value that you program with SENSe:CURRent:RANGe must be
the maximum current that you expect to measure. The instrument will select the range that gives the best
resolution. The crossover value is 20 mA. When queried, the returned value is the maximum current that
can be measured on the range that is presently set.
Command Syntax SENSe:CURRent[:DC]:RANGe[:UPPer] <NRf +>
Parameters 0 through MAX (see table 8-3)
Unit A (amperes)
*RST Value MAX (high range)
Examples SENS:CURR:RANG 4.0
Query Syntax SENSe:CURRent:RANGe?
Returned Parameters <NR3>