3 - Installation
44
Dc sources may be connected to the GPIB interface in series configuration, star configuration, or a
combination of the two, provided the following rules are observed:
The total number of devices including the GPIB interface card is no more than 15.
The total length of all cables used is no more than 2 meters times the number of devices connected
together, up to a maximum of 20 meters. (Refer to table 2-2 for a list of GPIB cables available from
Agilent Technologies.)
Do not stack more than three connector blocks together on any GPIB connector.
Make sure all connectors are fully seated and the lock screws are firmly finger-tightened.
RS-232 Interface
Agilent 66111A and 66311B/D dc sources have an RS-232 programming interface, which is activated by
commands located in the front panel Address menu. All applicable SCPI and COMPatibility commands
are available through RS-232 programming. When the RS-232 interface is selected, the GPIB interface is
disabled.
The RS-232 connector is a DB-9, male connector. Adapters are available to connect the dc source to any
computer or terminal with a properly configured DB-25 connector (see Table 2-2).
1 2 3 4 5
6 7 8 9
Figure 3-12. RS-232 Connector
Pin Input/Output Description
1 - no connection
2 Input Receive Data (RxD)
3 Output Transmit Data (TxD)
4 Output Data Terminal Ready (DTR)
5 Common Signal ground
6 Input Data Set Ready (DSR)
7 Output Request to Send (RQS)
8 Input Clear to Send (CTS)
9 - no connection