8 – Language Dictionar y
100
Calibration Commands
Calibration commands let you enable and disable the calibration mode, change the calibration password,
calibrate current and voltage programming, and store new calibration constants in nonvolatile memory.
NOTE: If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error. You must also save any changes that you made
using CALibrate:SAVE, otherwise all changes will be lost when you exit calibration mode.
CALibrate:CURRent
This command initiates the calibration of the positive dc output current as well as the high-range current
measurement circuit.
Command Syntax CALibrate:CURRent[:SOURce][:DC][:POSitive]
Parameters None
Examples CAL:STAT 1, ! enable calibration
CAL:CURR, ! start current calibration
Related Commands CAL: CURR:NEG, CAL:LEV, CAL:DATA
CALibrate:CURRent2
Agilent 66309B/D only
This command initiates the current calibration of output 2.
Command Syntax CALibrate:CURRent2
Parameters None
Examples CAL:CURR2
Related Commands CAL: CURR:NEG, CAL:LEV, CAL:DATA
CALibrate:CURRent:MEASure:LOWRange
Agilent 66311B/D, 66309B/D only
This command initiates the calibration of the low-range current measurement circuit.
Command Syntax CALibrate:CURRent:MEASure[:DC]:LOWRange
Parameters None
Examples CAL:CURR:MEAS
Related Commands CAL: CURR
CALibrate:CURRent:MEASure:AC
Agilent 66311B/D, 66309B/D only
This command initiates the calibration of the high bandwidth (ac) measurement circuit.
Command Syntax CALibrate:CURRent:MEASure:AC
Parameters None
Examples CAL:CURR:MEAS:AC