B - Verification and Calibration
152
Table B-2. Verification Test Record for Agilent Model 66111A
Model Agilent 66111A Report No.____________Date_____________
Test Description Minimum
Specification Recorded Results M aximum
Specification
Voltage Programming and Measurement
Low Voltage Vlo 10 mV _____mV +10 mV
Front Panel measurement Vfplo Vlo 5 mV _____mV Vlo +5 mV
High Voltage Vhi 14.982 V _______V 15.018 V
Front Panel measurement Vfphi Vhi 9.5 mV _______V Vhi +9.5 mV
Current Programming and Measurement
Low Current Ilo 1.33 mA _____mA 1.33 mA
High Current Ihi 2.9972 A _______A 3.0028 A
Front Panel measurement Ifphi Ihi 15 mA _______A Ihi +15 mA
Current Sink Measurement
Low Current Sink measurement Isinklo Isinklo 22.5 µA_____mA Isinklo +22.5 µA
High Current Sink measurement I sinkhi Isinklo 13 mA _____A Isinklo +13 mA
Table B-3. Verification Test Record for Agilent Model 66311B/D
Model Agilent___________ Report No.____________Date_____________
Test Description Minimum
Specification Recorded Results M aximum
Specification
Voltage Programming and Measurement
Low Voltage Vlo 10 mV _____mV +10 mV
Front Panel measurement Vfplo Vlo 5 mV _____mV Vlo +5 mV
High Voltage Vhi 14.982 V _______V 15.018 V
Front Panel measurement Vfphi Vhi 9.5 mV _______V Vhi +9.5 mV
Current Programming and Measurement (High Range)
Low Current Ilo 1.33 mA _____mA 1.33 mA
High Current Ihi 2.9972 A _______A 3.0028 A
Front Panel measurement Ifphi Ihi 6.5 mA _______A Ihi +6.5 mA
Current Measurement (Low Range)
Low Current measurement Ilrlo Ilrlo 2.5 µA _____µA Ilrlo +2.5 µA
High Current measurement Ilrhi Ilrhi 22.5 µA_____mA Ilrhi +22.5 µA
Current Sink Measurement
Low Current Sink measurement Isinklo Isinklo 22.5 µA_____mA Isinklo +22.5 µA
High Current Sink measurement I sinkhi Isinkhi 5.1 mA _____A Isinkhi +5.1 mA
DVM Voltage Measurement (Agilent 66311D only)
Positive Voltage measurement Dvpos Dvpos 15 mV _____V Dvpos +15 mV
Negative Voltage measurement Dvneg Dvneg 6.8 mV _____V Dvneg +6.8 mV