8 – Language Dictionar y
116
OUTPut:PROTection:CLEar
This command clears the latch that disables the output when an overvoltage, overcurrent,
overtemperature, or remote inhibit status condition is detected. All conditions that generate the fault must
be removed before the latch can be cleared. The output is then restored to the state it was in before the
fault condition occurred.
Command Syntax OUTPut:PROTection:CLEar
Parameters None
Examples OUTP:PROT:CLE
Related Commands OUT P:PROT:DEL *RCL *SAV
OUTPut:PROTection:DELay
This command sets the time between the programming of an output change that produces a constant
current condition (CC) and the recording of that condition by the Operation Status Condition register.
The delay prevents the momentary changes in status that can occur during reprogramming from being
registered as events by the status subsystem. Since the constant current condition is used to trigger
overcurrent protection (OCP), this command also delays OCP. Overvoltage protection is not affected by
this command.
Command Syntax OUTPut:PROTection:DELay <NRf+>
Parameters 0 to 2,147,483.647
Unit seconds
*RST Value 0.08 (Normal)
Examples OUTP:PROT:DEL 75E-1
Query Syntax OUTPut:PROTection:DELay?
Returned Parameters <NR3>
Related Commands OUT P:PROT:CLE CURR:PROT:STAT *RCL *SAV
OUTPut:RI:MODE
This command selects the mode of operation of the Remote Inhibit protection. The RI mode is stored in
non-volatile memory. The following modes can be selected:
LATChing causes a TTL low signal on the INH input to disable the output. The only way to clear
the latch is by sending OUTPut:PROTection:CLEAR while the INH input is false.
LIVE allows the INH input to disable the output in a non-latching manner. In other words,
the output follows the state of the INH input. When INH is low true, the output is
disabled. When INH is high the output is not affected.
OFF the INH input is disabled.
Command Syntax OUTPut:RI:MODE <mode>
Parameters LATChing | LIVE | OFF
Examples OUTP:RI:MODE LIVE
Query Syntax OUTPut:RI:MODE?
Returned Parameters <CRD>
Related Commands OUT P:PROT:CLE
OUTPut:TYPE
This command programs the output compensation circuit. This circuit compensates the output of the dc
source according to the input capacitance of the phone being tested. The circuit covers the following
capacitance ranges: LOW mode: 0 µF to 12,000 µF
HIGH mode: 5 µF to 12,000 µF
H2 mode: keeps the unit in High mode at all times