B - Verification and Calibration

DVM Calibration (applies to Agilent 66311D, 66309D only)

Action

Display

41.Connect the DVM inputs directly to output 1. Connect the external DMM to the DVM inputs as shown in figure B-1D. Do not connect the Agilent 3478 DMM.

42.

Press Shift Cal, scroll to CAL DVM, and press Enter.

CAL:DVM

43.

Press Shift Cal, scroll to CAL LEV, and press Enter to select the first

CAL:LEV P1

 

calibration point.

 

44.

Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter

CAL:DATA 0.00

 

the voltage value displayed on the external DMM. Numbers entered from

 

 

the external DMM should have an accuracy and resolution of 0.1 mV or

 

 

better.

 

45.Reverse the output 1 connections. Move the - lead to the + DVM connector pin, and move the + lead to the - DVM connector pin.

46.

Press Shift Cal, scroll to CAL LEV, use to scroll to P2 (the second

CAL:LEV P2

 

calibration point), and press Enter.

 

47.

Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter

CAL:DATA 0.00

the second voltage value displayed on the external DMM. Numbers entered from the external DMM should have an accuracy and resolution of 0.1 mV or better.

Because this second value is negative, you must place a minus sign in front of the value. Pressing the Enter Number key the second time enters the minus sign.

Saving the Calibration Constants

WARNING: Storing calibration constants overwrites the existing ones in non-volatile memory. If you are not sure you want to permanently store the new constants, omit this step. The dc source calibration will then remain unchanged.

 

Action

Display

48.

Press Shift Cal, scroll to CAL SAVE, and press Enter.

CAL:SAVE

49.

Press Shift Cal, select CAL OFF, and press Enter to exit Calibration mode.

CAL OFF

 

*RST and *RCL will also set the calibration state to OFF.

 

Restore the Language Setting

NOTE: Perform this procedure only if you changed the programming language from COMP to SCPI in the beginning of the calibration procedure.

 

Action

Normal Result

50.

Access the Address menu and scroll to the LANG: command.

LANG: SCPI

51.

Change the language from SCPI to COMPatibility, and press Enter.

LANG: COMP

158

Page 158
Image 158
Agilent Technologies 66309B DVM Calibration applies to Agilent 66311D, 66309D only, Saving the Calibration Constants, 158

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.