Table 8-1. Subsystem Commands Syntax (continued)

:DVM [:DC]?

Returns DVM dc voltage measurement

:ACDC?

Returns DVM rms voltage measurement

:VOLTage [:DC]?

Returns dc voltage

:ACDC?

Returns the total rms voltage (ac+dc)

:HIGH?

Returns the HIGH level of a voltage pulse

:LOW?

Returns the LOW level of a voltage pulse

:MAX?

Returns maximum voltage

:MIN?

Returns minimum voltage

OUTPut[12]

 

[:STATe] <bool>

Enables/disables the dc source output

:DFI

 

[:STATe] <bool>

Enables/disables the DFI output

:SOURce <source>

Selects event source (QUES OPER ESB RQS OFF)

:PON

 

:STATe <state>

Set power-on state (*RST RCL0)

:PROTection

 

:CLEar

Reset latched protection

:DELay <n>

Delay after programming/before protection

:RELay:

 

:MODE <mode>

Specifies the output relay mode (DD, HD, DH, or HH).

:RI

 

:MODE <mode>

Sets remote inhibit operating mode (LATC LIVE OFF)

:TYPE [:CAPacitance] <setting>

Sets output capacitor compensation (HIGH H2 LOW)

SENSe

 

:CURRent

 

[:DC]

 

:RANGe [:UPPer] <n>

Selects the high current measurement range

:DETector <detector>

Selects the current measurement detector (ACDC DC)

:FUNCtion <function>

Configures the measurement sensor ("VOLT" "CURR" "DVM")

:PROTection

 

:STATe <state>

Enables/disables open sense lead detection

:SWEep

 

:OFFSet

 

:POINts <n>

Defines the pre/post data capture in the measurement

:POINts <n>

Define the number of data points in the measurement

:TINTerval <n>

Sets the digitizer sample spacing

:WINDow [:TYPE] <type>

Sets the measurement window function (HANN RECT)

[SOURce:]

 

CURRent

 

[:LEVel]

 

[:IMMediate][:AMPLitude] <n>

Sets the output current limit

:TRIGgered [:AMPLitude] <n>

Sets the triggered output current limit

:PROTection

 

:STATe <bool>

Enables/disables current limit protection

CURRent2

 

[:LEVel]

 

[:IMMediate][:AMPLitude] <n>

Sets the output2 current level

:TRIGgered [:AMPLitude] <n>

Sets the triggered output2 current level

DIGital

 

:DATA [:VALue] <n>

Sets and reads the digital control port

:FUNCtion <function>

Configures digital control port (RIDF DIG)

VOLTage

 

[:LEVel]

 

[:IMMediate][:AMPLitude] <n>

Sets the output voltage level

:TRIGgered [:AMPLitude] <n>

Sets the triggered output voltage level

97

Page 97
Image 97
Agilent Technologies 66311B, 66309B, 66111A manual Language Dictionary

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.