Table 5 Troubleshooting Detector Issues (continued)

Problem

Possible Cause

Diagnosis

Corrective Action

 

 

 

 

Low Response

Contaminated ceramic

If there does not appear to

Replace tubes.

 

tubes.

be a leak, then the tubes

 

 

 

should be inspected.

 

 

 

Contamination can result

 

 

 

from column bleed, samples

 

 

 

which may contain volatile

 

 

 

metal complexes, and large

 

 

 

injections of coke forming

 

 

 

hydrocarbons.

 

Wandering Baseline

Poor temperature control.

Monitor the temperature on

 

 

the Controller. It should vary

 

 

by no more than ±5 °C.

Check for a loose connection on the heating element or thermocouple. Reposition the thermocouple in the Burner.

Wandering Baseline

Contamination in one of the

 

Detector gases.

Check the difference in the

Change Detector gases after

output signal between ozone

adding in-line traps.

on and off. The difference

 

should be between 0.2-2 mV

 

after equilibration.

 

Wandering Baseline

Leak in the oxidizer line at

 

the oxidizer inlet connection.

 

Leak at weld of H2 inlet

 

fitting.

If using SCD, use a microliter syringe containing a small amount of sulfur compound e.g., CS2 to “snoop” for leaks while watching mV output. Leaks are very evident by a large increase in signal displayed on the LED.

Reseat air line into air inlet nut, or replace.

Tailing Peaks with

Severe contamination of

High background signal with

Avoid “house” gases,

Non-Equimolar Response

Detector gases.

respect to ozone off.

especially air from

 

 

 

compressors. Clean gases

 

 

 

should be used with

 

 

 

appropriate traps.

 

 

 

 

Tailing Peaks

Poor column connection.

Verify column position at

Reinstall column.

 

 

inlet and outlet. Look for

 

 

 

discoloration of column at

 

 

 

Detector side which

 

 

 

indicates column in

 

 

 

combustion zone.

 

 

 

 

 

Tailing Peaks

Cracked tubes.

Confirm pressure and

Replace tubes as needed.

 

 

vacuum ranges. Inspect

 

 

 

columns and ferrules.

 

 

 

 

 

Controller Reads >1000 °C

Thermocouple open.

Check electrical resistance

Replace thermocouple.

 

 

between thermocouple pins.

 

 

 

 

 

114

Operation and Maintenance Manual

Page 114
Image 114
Agilent Technologies G6600-90006 manual Detector gases