Agilent Technologies G6600-90006 manual Reaction Cell, PMT Housing and PMT Socket

Models: G6600-90006

1 118
Download 118 pages 49.7 Kb
Page 86
Image 86

 

 

PMT SOCKET

PMT HOUSING

PMT

 

FILTER

 

 

 

HIGH

SIGNAL

 

VOLTAGE

 

 

to power supply

to amplifier

REACTION

 

PMT SOCKET

 

 

CELL

 

 

Figure 23 Reaction Cell, PMT Housing and PMT Socket

PRESSURE

SENSOR

O-RINGS

REDUCING

TRANSFERUNION LINE

FROM OZONE

GENERATOR

Figure 24 Reaction Cell

6Use a 7/64" Allen wrench to remove the three mounting screws from the reaction cell. Slowly pull the reaction cell back from the PMT housing. The optical filter is located between the reaction cell and the PMT housing. Remove the optical filter from the housing by tipping up the back of the Detector, if necessary, and allow the filter to fall onto a soft cloth.

7Inspect the o-rings and replace them if they show any wear.

86

Operation and Maintenance Manual

Page 86
Image 86
Agilent Technologies G6600-90006 manual Reaction Cell, PMT Housing and PMT Socket