Reaction cell pressure, with air to ozone generator OFF (0 psi setting on internal

regulator) and transfer line capped: ___________________ torr

Tighten connections if necessary and check to make sure pressure stabilizes in the expected region. If proper pressure is not obtained contact Agilent for assistance. If the reaction cell pressure is within the expected range, record the value, reset the internal air regulator to 3-6 psi, turn the pump OFF and proceed with the recorder test.

Recorder Test

The standard 355 SCD and 255 NCD recorder output configuration is 1 volt full scale. In addition, an output range can be set with the recorder output selection switch located on the back panel (Figure 11). Use a small screwdriver to adjust the switch to the proper position for your integrator or data system.

To check that the recorder cable has been properly connected to the integrator or data system, set the integrator to a high sensitivity setting (e.g. attenuation of x 1) and plot the background signal. Use the Output Offset to decrease the Detector baseline. If the recorder cable is connected correctly, the baseline will shift in response to changes in the Output Offset. If the polarity of the connection is incorrect, a negative response will be observed when the baseline is increased, and vice versa. Switch the polarity of the wires to correct this problem. If no response is observed, re-check the signal connections and repeat the test. Contact Agilent if there is no data system response after completion of this test.

68

Operation and Maintenance Manual

Page 68
Image 68
Agilent Technologies G6600-90006 manual Recorder Test