Agilent Technologies G6600-90006 manual Troubleshooting Burner Issues

Models: G6600-90006

1 118
Download 118 pages 49.7 Kb
Page 116
Image 116

Table 7 Troubleshooting Burner Issues

Problem

Possible Cause

Diagnosis

Corrective Action

Fuses Blow on Startup

Shorted heater element.

Inspect for exposed wires or

 

 

broken insulation around

 

 

wire leads. Measure

 

 

resistance leads, should not

 

 

be 0.

Take Burner top off and reseat insulation around heater leads.

Low Sensitivity

Leaks at ferrules or fittings.

For SCD, use a microliter

Inspect fittings for scoring

 

 

syringe containing a small

and replace if necessary.

 

 

amount of sulfur compound

Replace ferrules if worn.

 

 

e.g., CS2 to “snoop” for leaks

Hand tighten all fittings

 

 

while watching mV output.

mounted onto vespel

 

 

Leaks are very evident by a

ferrules, and wrench tighten

 

 

large increase in signal

all connections with Valco

 

 

displayed on the LED.

ferrules.

 

 

For NCD, a high background

 

 

 

could be observed from

 

 

 

atmospheric nitrogen.

 

 

 

 

 

Low Sensitivity

Improper gas flows.

Measure gas flows.

Reset gas flows.

 

 

 

 

Low Sensitivity

Broken ceramic tubes or

Remove and inspect tubes

Replace tubes or ferrules.

 

ferrules

and ferrules for chips and/or

 

High Pressure

 

breaks.

 

 

 

 

 

Low Sensitivity

Improper column

Inspect for charred Detector

After cutting off charred end

 

positioning.

end of column.

of column, measure column

High Background

 

 

length 108-109 mm from

 

 

 

ferrule (or 114-115 mm from

 

 

 

bottom of nut) and re-attach

 

 

 

column.

 

 

 

 

116

Operation and Maintenance Manual

Page 116
Image 116
Agilent Technologies G6600-90006 manual Troubleshooting Burner Issues