The baseline signal from the SCD and NCD can be adjusted from 0 to ±1% of the full scale recorder output using the offset control. (e.g., +10 mV to -10 mV for a 1 V full scale setting). This control can be used to offset the background signal.

Dual Plasma Controller Controls

The Agilent Dual Plasma Controller provides easy access to basic settings.

Figure 22 Dual Plasma Controller Front Panel

Oxidizer and

Allow you to adjust the oxidizer and hydrogen flow

Hydrogen Control

rate.

Knobs

 

Temperature Control

Allows you to adjust the Burner temperature.

Knob

 

Selector Control Knob

Changes the display to show the current value or

 

set point for each setting (temperature in °C,

 

pressure in Torr, oxidizer in sccm, or hydrogen in

 

sccm).