Repeatability*

<1.5% RSD 8 hours (~ the same in N and

 

nitrosamine)

 

<2% RSD 18 hours (~3% RSD in nitrosamine

 

mode over 21 hours)

Gas flow through Ozone Generator

20-30 mL/min at 3-6 psig (inlet pressure)

Typical reaction cell pressure

4 - 8 Torr RV5 Oil Sealed Pump

 

6 - 12 Torr Dry Piston Pump

Typical Burner Pressure

130 - 150 torr operating

Typical Burner Temperature

900 °C

Typical Hydrogen Flow Rate

6-10 SCCM

Typical Oxygen Flow Rate

8 - 15 SCCM

Signal Output Ranges

0-1V, 0-10V, 0 - 10V

*Burner temperature 950 °C, 11 SCCM oxygen, and 6 SCCM hydrogen; 25 ppm N as nitrobenzene in toluene; 0.2 µL injection on column (HP 19095-121Z), n=7 for 3 hours; n=13 for 18 hours and n=10

n-dipropylnitrosamine in toluene at 4 µg/mL, 0.2 µL injection on column (HP 19095-121Z).

Physical Specifications

Power requirements

355 SCD Detector

115 VAC, 50/60 Hz, 1400 W

 

100 VAC, 50/60 Hz, 1400 W

 

220-240 VAC, 50/60 Hz, 650W

255 NCD Detector

115 VAC, 50/60 Hz, 1400 W

 

100 VAC, 50/60 Hz, 1400 W

 

220-240 VAC, 50/60 Hz, 650W

Dual Plasma Controller

100-120 VAC, 50/60 Hz, 200 W

 

220-240 VAC, 50/60 Hz, 200 W

Dimensions and weight

Detector

Height: 16.0 in (40.6 cm)

 

Width: 9.2 in (23.4 cm)

 

Depth: 21.8 in (55.3 cm)

 

355

SCD Weight: 34.0 lbs (15.0 kg)

 

255

NCD Weight: 37.5 lbs (17.0 kg)