Step 5: Install the Dry Compressed Air or O2 Supply

Connect a 1/8" OD Teflon (PFA) line fitted with a 1/8" brass Swagelok nut from the AIR INLET at the rear of the Detector (see Figure 17) to a supply of dry compressed air or oxygen. The air regulator located inside the front door of the Detector should be set to approximately 3-6 psi.