Agilent Technologies G6600-90006 manual Detector Front Panel Controls, Power Controls

Models: G6600-90006

1 118
Download 118 pages 49.7 Kb
Page 64
Image 64

Detector Front Panel Controls

As illustrated in Figure 21, the front panel is divided into three sections: signal control, display output control, and power control. Each section is described below. All the front panel LEDs are red in the ON mode and darkened in the OFF mode except for POWER. This LED toggles green for ON and red for STANDBY. In STANDBY mode, the front panel display and high voltage to the photomultiplier tube (PMT) are turned off. The red LED in the STANDBY mode serves as a reminder that power is still supplied to the instrument.

Figure 21 Front Panel Controls

Power Controls

ON: Turns on front panel displays and high voltage to the PMT.

STANDBY: Darkens panel and turns off high voltage to the PMT. Does not shut off main power.

Turns pump ON (red) and OFF (dark).

Functions in both ON and STANDBY modes.

64

Operation and Maintenance Manual

Page 64
Image 64