Agilent Technologies G6600-90006 manual Detection Limits

Models: G6600-90006

1 118
Download 118 pages 49.7 Kb
Page 76
Image 76

Detection Limits

The following table lists the detection limits which can be expected for typical chromatographic conditions, assuming proper operation of the Detector and chromatographic systems.

Table 2 Expected Detection Limits for Chromatographic Conditions

Type of Injection

Volume

Column

Detection Limit Per

 

 

 

 

Compound as Sulfur

 

 

 

 

 

Liquid, Split 1:10

1

µL

Capillary 0.32 mm ID

50 parts per billion

 

 

 

 

 

Liquid, Split 1:100

1

µL

Capillary 0.32 mm ID

0.5 parts per million

 

 

 

 

 

Liquid, On-column or splitless

1

µL

Capillary 0.32 mm ID

5 parts per billion

 

 

 

 

 

Liquid, Splitless

10

µL

Capillary 0.32 mm ID

0.5 parts per billion

 

 

 

 

Gas, Split 1:10

1 cm3

Capillary 0.32 mm ID

50 parts per billion

Gas, Direct on-column

1 cm3

0.53 mm ID

5 parts per billion

Gas, Direct on-column

1 cm3

Packed

5 parts per billion

Gas, Direct on-column

10 cm3

Packed

0.5 parts per billion

76

Operation and Maintenance Manual

Page 76
Image 76
Agilent Technologies G6600-90006 manual Expected Detection Limits for Chromatographic Conditions