Detection Limits
The following table lists the detection limits which can be expected for typical chromatographic conditions, assuming proper operation of the Detector and chromatographic systems.
Table 2 Expected Detection Limits for Chromatographic Conditions
Type of Injection | Volume | Column | Detection Limit Per | |
|
|
|
| Compound as Sulfur |
|
|
|
|
|
Liquid, Split 1:10 | 1 | µL | Capillary ≤0.32 mm ID | 50 parts per billion |
|
|
|
|
|
Liquid, Split 1:100 | 1 | µL | Capillary ≤0.32 mm ID | 0.5 parts per million |
|
|
|
|
|
Liquid, | 1 | µL | Capillary ≤0.32 mm ID | 5 parts per billion |
|
|
|
|
|
Liquid, Splitless | 10 | µL | Capillary ≤0.32 mm ID | 0.5 parts per billion |
|
|
|
| |
Gas, Split 1:10 | 1 cm3 | Capillary ≤0.32 mm ID | 50 parts per billion | |
Gas, Direct | 1 cm3 | 0.53 mm ID | 5 parts per billion | |
Gas, Direct | 1 cm3 | Packed | 5 parts per billion | |
Gas, Direct | 10 cm3 | Packed | 0.5 parts per billion |
76 | Operation and Maintenance Manual |