Detection Limits

The following table lists the detection limits which can be expected for typical chromatographic conditions, assuming proper operation of the Detector and chromatographic systems.

Table 2 Expected Detection Limits for Chromatographic Conditions

Type of Injection

Volume

Column

Detection Limit Per

 

 

 

 

Compound as Sulfur

 

 

 

 

 

Liquid, Split 1:10

1

µL

Capillary 0.32 mm ID

50 parts per billion

 

 

 

 

 

Liquid, Split 1:100

1

µL

Capillary 0.32 mm ID

0.5 parts per million

 

 

 

 

 

Liquid, On-column or splitless

1

µL

Capillary 0.32 mm ID

5 parts per billion

 

 

 

 

 

Liquid, Splitless

10

µL

Capillary 0.32 mm ID

0.5 parts per billion

 

 

 

 

Gas, Split 1:10

1 cm3

Capillary 0.32 mm ID

50 parts per billion

Gas, Direct on-column

1 cm3

0.53 mm ID

5 parts per billion

Gas, Direct on-column

1 cm3

Packed

5 parts per billion

Gas, Direct on-column

10 cm3

Packed

0.5 parts per billion

76

Operation and Maintenance Manual

Page 76
Image 76
Agilent Technologies G6600-90006 manual Expected Detection Limits for Chromatographic Conditions