Note the position of the oil level in the window after operating the pump for several hours. For the next several days of operation, recheck the oil level daily. If the oil level is increasing, water is accumulating in the oil reservoir. Confirm that the water vapor is properly being expelled from the oil mist filter. If the oil level is dropping, excess air flow through the pump is forcing oil out of the vacuum pump. Turn the restrictor plate so that the single indentation on the restrictor plate is aligned with the indentation on the side of the oil return assembly. After adjustment of the gas ballast, allow the system to operate for an additional day and check the oil level again.

It is normal for the oil to appear foamy from air in the oil when viewing the oil in the oil level window. The purpose of the ballast control is to sweep ambient air into the pump oil. Ensure the oil level when the pump is operating is not above the “Full” mark on the pump.