Specifications

Technical Information — 355 SCD

Sensitivity*

Typical Selectivity

Linearity

Precision and Stability

Ozone flow through the Post Ozone Restrictor Typical reaction cell pressure

Typical Burner Pressure

Typical Burner Temperature

Typical Air Flow Rate

Typical Hydrogen Flow Rate

Signal Output Ranges

Typical time to reach 800 °C from room temperature

Typical safety shroud outside temperature Recorder output

Typical < 0.5 pg S/second (signal to noise 3.3:1) g S/g C > 2 x 107

>104

<2% RSD 2 hours <5% RSD 72 hours

20-30 mL/min at 3-6 psig

4 - 8 Torr RV5 Oil Sealed Pump

6 - 12 Torr Dry Piston Pump

250-400 Torr operating

800 °C

65 SCCM recommended

3-10 SCCM recommended with FID adapter

40 SCCM recommended

0-1V, 0-10V

10 min typical(120 VAC, 60 Hz)

<65 °C at 800 °C Burner temperature typical 0-1 V or 0-10 V

* Burner temperature 800 °C, 80 SCCM air, and 60 SCCM hyrdrogen, test compound dimethyl sulfide in toluene.

Based on thiopene in benzene at 1 ppm mass sulfur, 1 µL injection split 1:10, 30 m, 0.32 mm ID, 1 µm thick CP wax (n=10 for 2.5 hours; n=42 for 72 hours).

Subject to change without notice.

Technical Information — 255 NCD

Sensitivity

<3 pg N/second (signal to noise 3:1) in both N

 

and nitrosamine modes

Selectivity

g N/g C > 2 to 107 in N mode (selectivity in

 

nitrosamine mode is matrix-dependent)

Linearity

>104

20