Agilent Technologies G6600-90006 manual Troubleshooting Pump Issues

Models: G6600-90006

1 118
Download 118 pages 49.7 Kb
Page 115
Image 115

Table 5 Troubleshooting Detector Issues (continued)

Problem

Possible Cause

Diagnosis

Corrective Action

 

 

 

 

Burner Fault Cycles. (off and

Cracked tube.

Pressure increases

Replace broken tubes.

ON)

 

>525 torr, Burner faults.

 

 

 

Cycle repeated as

 

 

 

temperature and pressure

 

 

 

reset.

 

 

 

 

 

Burner Fault Cycles (off and

Leak in Burner.

Pressure increases

Locate and repair leaks,

ON)

 

>600 torr, Burner faults.

check integrity of ferrules.

 

 

Cycle repeated as

 

 

 

temperature and pressure

 

 

 

reset.

 

 

 

 

 

Table 6 Troubleshooting Pump Issues

Problem

Possible Cause

Diagnosis

Corrective Action

Pump is noisy

High-pressure dry piston.

Pressure in the reaction cell Change sleeve/seal. is unacceptably high and the

pump is getting audibly noisy.

Pump Doesn’t Start

Dry piston seals have

Pump fuses blow

Dry piston requires break-in

 

recently been changed.

immediately after changing

time of a few minutes after

 

 

the seals and sleeves.

seals are changed.

 

 

 

 

Pump Doesn't Start

Pump switch off (dry piston

Locate pump switch on

Turn ON switch.

 

pump must start from

pump and verify position.

 

 

atmospheric pressure).

 

 

 

 

 

 

Fuses Blow on Startup

Emulsified oil (worn seals

Inspect oil for integrity.

Change pump oil, and plug

 

and head on dry piston

 

unit into wall to run for 10-15

 

pump).

 

minutes.

 

 

 

 

Water in Pump

Cracked coalescing filter.

Milky yellow oil in the pump

Change coalescing filter and

 

 

window.

pump oil.

 

 

 

 

Reaction Cell Pressure High

chemical trap clogged.

Remove trap from the

Change chemical trap.

 

 

vacuum line and confirm

 

 

 

expected pressure readings.

 

 

 

 

 

Reaction Cell Pressure High

Leak in ozone generator.

Air regulator falls rapidly to

Replace ozone generator.

 

 

zero when turned off.

 

 

 

 

 

Pump Loses Oil

Ballast Open.

Oil level drops.

Reset ballast

Gurgle Sound

 

 

See Pump specific sections.

 

 

 

 

High level of oil in

Plugged oil return restrictor

No visible movement of oil in

Change filter and clear

Coalescing Filter

 

the return line.

restrictor.

 

 

 

 

Operation and Maintenance Manual

115

Page 115
Image 115
Agilent Technologies G6600-90006 manual Troubleshooting Pump Issues