Step 7: Install the Dual Plasma Burner

Remove the cover plates from the Detector area of the GC to expose the hole into the oven through which a Detector is normally mounted. If the GC has more than one available Detector position, pick the most convenient one.

Prepare the GC by cutting the inside liner and top liner per Figure 17. Note that mounting fastener patterns will vary by GC manufacturer. Make sure the notch in the inside liner is on the right, in order to accommodate the geometry of the Burner as it sits in the shroud and mounting plate.

Figure 18 Dimensions of GC Liner Cut-Outs

 

Align the Burner mounting plate with the mounting screw holes on the GC.

 

Clear the hole into the GC oven of interfering insulation, and then secure the

 

mounting plate onto the oven with the screws provided. Attach the Burner's

 

heated base connector to the GC's temperature control circuit. Consult your

 

GC's operation or service manual to confirm proper connection of the 100 ohm

 

RTD sensor and the cartridge heaters.

 

Occasionally geometric design changes will occur within or between GC models. If the

N O T E

mounting plate provided does not match up with the top of the GC, contact Agilent for

 

 

additional information.

 

 

 

Position the Dual Plasma Burner (column end down) into the tapered fitting of

 

the heated base, with the lower hydrogen line and pin aligned with the slot in

 

the heated base. The Burner should be secure when properly positioned.