Agilent Technologies G6600-90006 manual Two-stage pressure regulators

Models: G6600-90006

1 118
Download 118 pages 49.7 Kb
Page 43
Image 43

with air as the oxidant in the SCD and oxygen as the oxidant in the NCD.

For the SCD, detector gases must be sulfur free (<1ppb) for proper Detector operation. In general, bottled air is preferred to “house” air from a compressor because compressors tend to generate lower quality air. In addition, pressure fluctuations induced by the air generator may be detrimental to Detector performance.

For the NCD and GC, use gases that are low in nitrogen and total hydrocarbons. Use gases with an "instrument" or "chromatographic" purity rating. Agilent recommends gases with a purity range of 99.995% to 99.9995%. Do not use oil-pumped supplied air since it contains large concentrations of hydrocarbons.

Use of appropriate traps on carrier, hydrogen and oxidant sources to improve support gas quality, such as a sulfur trap gas purifier, is recommended. The oxidant used to supply the ozone generator should be dry to prevent internal corrosion. Use of a moisture trap or dryer is recommended.

Two-stage pressure regulators

The combustion gases must be supplied to the Controller at a pressure of

25 psig or less. Use a two-stage pressure regulator rather than a single-stage regulator to eliminate pressure surges. High-quality, stainless-steel diaphragm-type regulators are recommended. On/Off valves at the regulators are useful but are not essential. Mount the valves at the outlet fitting of the pressure regulators.

If you use gases with a two-stage regulator up-stream, it will be sufficient to use single-stage regulators at the Detector.

Particle Filtration

Up-stream, inline particle filtration of 10 micron or better is required to prevent damage to Controller and Burner components.

WARNING

Ozone is a hazardous gas and a strong oxidant. Minimize exposure to ozone by using the

instrument in a well-ventilated area and venting the exhaust of the vacuum pump to a

 

 

fume hood. Turn off the ozone generator when the instrument is not in use.

 

 

 

Supply Tubing for Combustion Gases

 

Use preconditioned and cleaned copper or stainless steel tubing to supply

 

gases to the 355 SCD and 255 NCD. Do not use ordinary copper tubing since it

 

contains oils and other contaminants. Plastic tubing is not recommended since

 

it is permeable to oxygen and other contaminants that can damage columns

 

and detectors or cause elevated background. Teflon tubing may be acceptable

 

in some clean environments.

Operation and Maintenance Manual

43

Page 43
Image 43
Agilent Technologies G6600-90006 manual Two-stage pressure regulators