Monitoring Oxidizer and Hydrogen Flow with the Dual Plasma Controller

Hydrogen will flow to the Burner only if the temperature is above 325 °C and the pressure of the Burner is <575 torr.

Turn ON the vacuum pump at the Detector and the power to the Dual Plasma Controller. The power LED should illuminate. When the temperature reaches approximately 325 °C, the valve’s LED should illuminate, indicating hydrogen and oxidant flow.

Use the Controller display knobs to select display of oxidizer or hydrogen flow.

WA R N I N G

Hydrogen gas is explosive and must be handled carefully. Keep away from sources of

ignition.

 

WA R N I N G

The Burner is hot. Do not touch the Burner. Let the Burner cool before performing any

operations involving the Burner.

 

70

Operation and Maintenance Manual

Page 70