Agilent Technologies G6600-90006 manual Operation and Maintenance Manual

Models: G6600-90006

1 118
Download 118 pages 49.7 Kb
Page 79
Image 79

can be achieved by eliminating the lower plasma. This is accomplished by plugging the side port of the splitter fitting and teeing the air into the lower hydrogen line, much like the configuration used for Nitrosamine analysis as shown in Figure 9. A slightly lower air flow rate, such as 40 SCCM and a hydrogen flow rate of around 60 SCCM is used. Sensitivity of around 0.1 pg S/sec or less should be readily achievable. It is not possible to cover all potential applications, so some optimization may be desirable.

Operation and Maintenance Manual

79

Page 79
Image 79
Agilent Technologies G6600-90006 manual Operation and Maintenance Manual