Instrument Shut-Down

Daily Shutdown

1Toggle off the ozone generator.

2Turn off the air regulator (counter-clockwise), located inside Detector door.

3Toggle power to "stand-by."

4Leave the vacuum pump and Dual Plasma Controller operating at all times.

Complete Shutdown

1Toggle off the ozone generator.

2Turn off the air regulator (counter-clockwise), located inside Detector door.

3Toggle power to STANDBY.

4Turn off power to the Dual Plasma Controller.

5After 15 minutes, toggle off the vacuum pump, so the Burner cools and moisture is removed from the system.

6Turn off gases to unit.

Operation and Maintenance Manual