Agilent Technologies G6600-90006 manual Plugged Restrictor Lines

Models: G6600-90006

1 118
Download 118 pages 49.7 Kb
Page 110
Image 110

Plugged Restrictor Lines

Plugged restrictor lines are verified by turning off the air/oxygen regulator located inside the door of the Detector and observing little or no change in the pressure as indicated on the regulator. The plug can be located in either the pre-ozone restrictor or the post-ozone restrictor (see Figure 3). A partially plugged restrictor line can result in a non-linear signal, or reduced sensitivity which will vary with the flow of ozone. Replace the restrictor lines.

110

Operation and Maintenance Manual

Page 110
Image 110
Agilent Technologies G6600-90006 manual Plugged Restrictor Lines