Diagnosing General Problems

Table 5 Troubleshooting Detector Issues

Problem

Possible Cause

Diagnosis

Corrective Action

 

 

 

 

No Response

No ozone

Little or no difference in

Continue below. See No

 

 

output signal between ozone

Ozone in Problem column.

 

 

ON vs. OFF. Typically the

 

 

 

background signal will be

 

 

 

0.3-0.8 mV higher with the

 

 

 

ozone ON.

 

 

 

 

 

No Ozone

Blown fuse

Ozone indicator light

Replace ozone fuse.

 

 

remains off when ozone

 

 

 

button is pushed.

 

No Ozone

High Voltage Transformer

No difference in output

 

and/or ozone generator is

signal between ozone ON vs.

 

inoperative.

OFF even though flow

 

 

through the ozone generator

 

 

is good (about

 

 

25-35 mL/min).

Have ONLY someone with high voltage experience use a high voltage probe to check the input to the ozone generator across the two wires. It should be

>6000 volts. If it is not then replace high voltage transformer. If it is, replace ozone generator.

No Ozone

Ozone restrictor(s) plugged.

Needle on the regulator

Replace plugged

 

 

inside front panel does not

restrictor(s).

 

 

move downward when

 

 

 

regulator is turned fully

 

 

 

counterclockwise. The

 

 

 

measured flow is low or

 

 

 

non-existent.

 

 

 

 

 

No Response

Hydrogen and/or air ran out.

Measure flow rates.

Set correctly.

 

 

 

 

No Response

Broken ceramic tube(s).

Pressure on Controller

Change tubes.

 

 

and/or Detector is too high

 

 

 

or too low.

 

 

 

 

 

No Response

Heating element failure.

Temperature at Controller

Replace heating element.

 

 

<100 °C.

 

 

 

 

 

Low Response

Improper hydrogen/air flow

Measure flow rates.

Adjust flow rates.

 

rates.

 

 

 

 

 

 

Low Response

Detector leaks.

Note pressures on

Locate and repair leaks,

 

 

Controller.

check integrity of ferrules.

 

 

 

 