15Carefully bend the H2 line into position so that the 1/16" Valco nut and ferrule can be screwed into the side port of the splitter fitting. Tighten the

connection of the H2 line to the splitter fitting using a 3/8" wrench on the vertical flat of the splitter fitting and 1/4" wrench on the Valco nut.

16Make sure that no other connections have loosened or moved out of alignment, if so, reposition or retighten the fittings as needed.

17The assembled Burner is now ready for re-installation on the GC.