Agilent Technologies G6600-90006 manual Ozone Generator

Models: G6600-90006

1 118
Download 118 pages 49.7 Kb
Page 28
Image 28

temperature, hydrogen and oxidant flow rates, and Burner pressure. The temperature, actual pressure, oxidant and hydrogen flow rates are selected for display by rotation of a 4-position control knob. Power, valve operation, temperature within set-point range and fault conditions are indicated with LED illumination on the front display panel.

The Dual Plasma Controller incorporates several safety features. The safety circuitry detects faults such as power loss, vacuum loss, thermocouple failure, heater element failure, broken ceramic tube, or high temperature. When a fault is detected, the Fault LED illuminates and hydrogen and oxidant flow is stopped by normally-closed solenoid valves.

Ozone Generator

The SCD and NCD produce ozone by corona discharge using a clean, pressurized air or oxygen source. Use of oxygen should increase ozone production and, hence, Detector response. High voltage to the ozone generator is applied only when the reaction cell pressure is less than 100 torr in the SCD and less than 200 torr in the NCD. Gas flow through the ozone generator is controlled by a pressure regulator and flow restrictors.

WA R N I N G

Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized by

using the instrument in a well ventilated area, changing the chemical trap regularly, and

 

 

venting the exhaust of the vacuum pump. The ozone generator should be turned off when

 

the instrument is not in use to reduce maintenance requirements.

 

 

Chemiluminescence Reaction Cell and Photomultiplier Tube

Sulfur monoxide (formed in the Burner) and ozone (produced in the ozone generator located in the Detector) are mixed in the reaction cell. The cell is designed such that the reaction between SO and O3 occurs directly in front of the photomultiplier tube (PMT). A UV band pass filter (300 - 400 nm) located between the reaction cell and the PMT selectively transmits the light emitted by the SO/O3 reaction. Efficient combustion in the ceramic tubes coupled with the UV band pass filter eliminates interference from non-sulfur containing analytes (e.g. nitric oxide, olefins, etc.) which also undergo chemiluminescent reactions with ozone. A background signal is typically present as a result of ozone-wall interactions and low level sulfur contamination of Detector gases. This background signal can be used as a troubleshooting aid (see Section 10).

28

Operation and Maintenance Manual

Page 28
Image 28
Agilent Technologies G6600-90006 manual Ozone Generator, Chemiluminescence Reaction Cell and Photomultiplier Tube