Anderson Manufacturing AV-9000 manual 11-3

Models: AV-9000

1 321
Download 321 pages 29.81 Kb
Page 228
Image 228

Step 4

Be sure that the jumpers on the Input Board for the Input being calibrated are positioned as shown in Table 11-3, and Figure 11-1 or Table 11-3A and Figure 11-1A, depending on which circuit board is fitted.

TABLE 11-3 INPUT BOARD JUMPER POSITIONS

 

 

 

INPUT 1, 3, 5, 7

 

 

 

 

INPUT 2, 4, 6, 8

 

 

 

 

 

 

 

 

 

 

 

 

JU1

JU2

JU3

JU7

JU11

JU15

JU4

JU5

JU6

JU8

JU12

JU16

TC NARROW

U

L

P

R

R

P

U

R

P

R

L

P

TC WIDE

U

L

P

M

R

P

U

R

P

M

L

P

RTD

P

L

U

L

L

P

P

R

U

L

R

P

mA

U

L

P

L

R

U

U

R

P

L

L

U

25 mV

U

L

P

R

R

P

U

R

P

R

L

P

100 mV

U

L

P

M

R

P

U

R

P

M

L

P

1 VOLT

U

L

P

L

R

P

U

R

P

L

L

P

10 VOLT

U

L

D

L

R

P

U

R

D

L

L

P

 

 

 

INPUT 1, 3, 5, 7

 

 

 

 

INPUT 2, 4, 6, 8

 

 

 

 

 

 

 

 

 

 

 

 

 

JU2

JU3

JU7

JU15

 

 

JU5

JU6

JU8

JU16

 

TC NARROW

 

P

P

R

P

 

 

P

P

R

P

 

TC WIDE

 

P

P

M

P

 

 

P

P

M

P

 

RTD

 

P

U

L

P

 

 

P

U

L

P

 

mA

 

P

P

L

F

 

 

P

P

L

F

 

25 mV

 

P

P

R

P

 

 

P

P

R

P

 

100 mV

 

P

P

M

P

 

 

P

P

M

P

 

1 VOLT P

 

P

P

L

P

 

 

P

P

L

P

 

10 VOLT

 

P

D

L

P

 

 

P

D

L

P

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CODE: D - DOWN, L - LEFT, M - MIDDLE, P - PARKED, R - RIGHT, U - UP, F - FITTED

FIGURE 11-1

 

JU15

 

JU16

 

 

JU7

JU3 JU1

 

JU4

JU6

JU8

 

 

 

JU11

JU2

 

 

JU5 JU12

 

- +

 

+ -

 

 

TB1

1

U6

1

 

TB2

 

 

JU7

JU3

JU2

JU15

- +

TB1 1

U1

JU8

 

JU6

 

JU5

 

JU16

+

-

1 TB2

There are up to an additional 2 jumpers per Input Board that are used for BOARD ID. Location (Board 1 is the bottom board) determines the position of JU13 and/or JU14.

Section 11

11-3

Edition 3

Rev G

Page 228
Image 228
Anderson Manufacturing AV-9000 manual 11-3