ACCURACY TABLES CONT.

To achieve stated results, the following thermocouples must be used with the INPUT TYPE/RANGE set to TC NARROW

 

See Note:

1

2

3

4

5

6

TC

RANGE

REF

LIN.

FACTORY

REF+LIN

DEVIATION

RESOL

TYPE

 

ACC'Y

ACC'Y

CAL

+CAL

ACC'Y

 

 

°C

°C

°C

°C

°C

°C

°C/bit

T

0/400

0.12

0.09

0.34

0.55

0.04

0.018

 

-200/0

0.23

0.19

0.70

1.12

0.02

0.036

 

-250/-220

0.73

0.36

1.90

2.98

0.00

0.111

R

800/1700

0.43

0.19

0.83

1.45

0.09

0.065

 

200/800

0.58

0.25

1.08

1.90

0.06

0.088

S

250/1750

0.56

0.25

1.05

1.87

0.15

0.086

B

200/1800

0.74

0.31

1.34

2.39

0.16

0.113

 

 

 

 

 

 

 

 

 

See Note:

1

2

3

4

5

6

RTD

RANGE

REF

LIN.

FACTORY

REF+LIN

DEVIATION

RESOL

TYPE

°C

ACC'Y

ACC'Y

CAL

+CAL

ACC'Y

°C/bit

 

°C

°C

°C

°C

°C

 

 

 

 

 

 

 

 

385

-160/480

0.16

0.03

0.13

0.33

0.06

0.025

DIN

-200/-160

0.14

0.20

0.12

0.46

0.00

0.022

392

-100/450

0.16

0.03

0.13

0.32

0.05

0.025

USA

 

 

 

 

 

 

 

392

-200/560

0.29

0.13

0.24

0.66

0.06

0.044

SAMA

 

 

 

 

 

 

 

100 ohm

-40/200

0.09

0.05

0.07

0.21

0.02

0.013

Nickel

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Edition 3

1-8

Section 1

Rev G

Page 16
Image 16
Anderson Manufacturing AV-9000 manual Sama