GE 90-30/20/Micro manual Setting User Parameters Including Tuning Loop Gains

Models: 90-30/20/Micro

1 311
Download 311 pages 52.91 Kb
Page 263
Image 263

4

Plotting a step response at time t0 in the time domain provides an open loop unit reaction curve:

CV Unit Step Output to Process

PV Unit Reaction Curve Input from Process

a45709

1

t0

K

 

0.632K

 

t0

 

Tp

Tc

The following process model parameters can be determined from the PV unit reaction curve:

KProcess open loop gain = final change in PV/change in CV at time t0 (Note no subscript on K)

Tp

Process or pipeline time delay or dead time after t0 before the process output PV

 

starts moving

 

 

Tc

First order Process time constant, time required after Tp for PV to reach 63.2% of the

 

final PV

Usually the quickest way to measure these parameters is by putting the PID block in Manual mode and making a small step in CV output, by changing the Manual Command %Ref+13, and plotting the PV response over time. For slow processes, this can be done manually, but for faster processes a chart recorder or computer graphic data logging package will help. The CV step size should be large enough to cause an observable change in PV, but not so large that it disrupts the process being measured. A good size may be from 2 to 10% of the difference between the CV Upper and CV Lower Clamp values .

Setting User Parameters Including Tuning Loop Gains

As all PID parameters are totally dependent on the process being controlled, there are no predetermined values that will work, however, it is usually a simple, iterative procedure to find acceptable loop gain.

1.Set all the User Parameters to 0, then set the CV Upper and CV Lower Clamps to the highest and lowest CV expected. Set the Sample Period to the estimated process time constant (above)/10 to 100.

2.Put block in Manual mode and set Manual Command (%Ref+13) at different values to check if CV can be moved to Upper and Lower Clamp. Record PV value at some CV point and load it into SP.

3.Set a small gain, such as 100 * Maximum CV/Maximum PV, into Kp and turn off Manual mode. Step SP by 2 to 10% of the Maximum PV range and observe PV response. Increase Kp if PV step response is too slow or reduce Kp if PV overshoots and oscillates without reaching a steady value.

4.Once a Kp is found, start increasing Ki to get overshooting that dampens out to a steady value in 2 to 3 cycles. This may required reducing Kp. Also try different step sizes and CV operating points.

4-182

Series 90-30/20/Micro Programmable Controllers Reference Manual–September 1998

GFK-0467K

Page 263
Image 263
GE 90-30/20/Micro manual Setting User Parameters Including Tuning Loop Gains