Error Messages

Appendix D

 

 

 

Error Messages

 

Error Messages

 

 

Error Message

Cause/Possible Action

 

 

Adjust volume too small

In DISPENSE, you have, you have entered a value for “Final

 

vol.” that is too small.

 

“Final vol.” > volume of biggest fraction dispensed.

 

 

Asp flow missing

Re-enter Method, supplying the missing parameter, when

 

prompted to do so (see DILUTE, DISPENSE, PIPETTE, and

 

MIX tasks).

 

 

Diluent volume missing

Re-enter Method, supplying the missing parameter, when

 

prompted to do so (see DILUTE task).

 

 

Disp flow missing

Re-enter Method, supplying the missing parameter, when

 

prompted to do so (see DILUTE, DISPENSE, PIPETTE, and

 

MIX tasks).

 

 

DISP task in wrong place

Within a Method, DISPENSE must follow DILUTE or MIX.

 

To DISPENSE before DILUTE you should create a Method file

 

containing only the DISPENSE task.

 

 

Incorrect flow rates

Use flow rates that correspond to the configured syringe

 

size(s); refer to page 5-31.

 

 

Left syr life exhausted!

For the Dual-syringe model: piston of left-hand syringe may

 

need replacing.

 

 

LINK method not allowed!

You can’t link files that are already linked.

 

 

Max sample vol. reached

In DILUTE UNKNOWN, you need a larger syringe/transfer-

 

tubing to measure volumes greater than the maximum

 

reached with the existing configuration.

 

 

Mechanical problem: Help

Occurs when the valve or a syringe jams; switch off and

 

correct the problem:

 

the syringe and/or valve need servicing (clean or change).

 

 

Method does not exist!

First, you must create the Method to be linked, then enter its

 

file number using LINK.

 

 

Method used in link file

You have attempted to delete a Method number that has been

 

assigned in LINK. You must delete the file containing LINK,

 

before deleting the Method.

 

 

Method xxx is locked

Unlock Method file, to continue with this action.

 

 

Mix volume is missing

Re-enter Method, supplying the missing parameter, when

 

prompted to do so (see MIX task).

 

 

Nb mixes is missing

Re-enter Method, supplying the missing parameter, when

 

prompted to do so (see MIX task).

 

 

No room to copy meth xxx

You need to free up space by deleting one or more Methods.

 

 

Not enough memory!

Reduce the size and/or number of stored Method files.

 

 

Appendix D-2

Page 116
Image 116
Gilson 402 manual Error Messages Appendix D

402 specifications

The Gilson 402 is a highly regarded laboratory device known for its precision and reliability in liquid handling applications. Primarily utilized in research and clinical labs, this automated pipetting system significantly enhances productivity by reducing the time taken for repetitive tasks while ensuring maximum accuracy.

One of the standout features of the Gilson 402 is its intuitive user interface. The touchscreen display allows operators to easily navigate through various functions and settings, enabling quick setup and adjustments. Its intuitive design minimizes the learning curve for new users, making it accessible even for those with limited technical expertise.

The Gilson 402 employs advanced pipetting technologies, which provide exceptional reproducibility and reliability. The system is equipped with proprietary Pipette G (as used in Gilson's PIPETMAN series), which ensures precise volume delivery. This is crucial in applications such as molecular biology, where even minor discrepancies in liquid volumes can lead to significant variations in experimental outcomes.

Another key characteristic of the Gilson 402 is its versatility. The device accommodates a wide range of pipette tips, allowing users to work with varying volumes and types of samples. Whether handling delicate biological fluids or viscous solutions, the Gilson 402 adapts to different protocols seamlessly.

The automation capabilities extend to a customizable programming feature that permits users to create complex pipetting protocols with ease. Multiple-step pipetting tasks can be completed effortlessly, reducing the likelihood of errors caused by manual handling.

Equipped with high-precision motors and sensors, the Gilson 402 also features built-in calibration and quality control functions. This guarantees that the system remains accurate over time and can easily be adjusted to maintain compliance with stringent laboratory standards.

In summary, the Gilson 402 is a cutting-edge liquid handling solution that combines advanced technology with user-friendly features. Its reliability, precision, and versatility make it an indispensable tool in modern laboratories. Whether used for routine tasks or complex experiments, the Gilson 402 enhances efficiency, promotes accuracy, and ultimately supports groundbreaking scientific research.