Gilson 402 Prefabricated Methods, Key to Press, Edit More Link File # 7 ENTER Next File # 8 ENTER

Models: 402

1 124
Download 124 pages 13.02 Kb
Page 78
Image 78
Key to Press:

Operation - Advanced Use

5

 

 

 

Software

Key in the file number of the second Method to link(8), press ENTER, press Next.

Key in the file number of the third Method to link(9), press ENTER, press Next.

Key in the file number of the fourth Method to link(8), press ENTER, press Next.

After a file is linked in this manner it can not be deleted.

Press ESC to access the Save Screen, press Yes.

Typical Task Entry

Key to Press:

File # (44) ENTER

Edit

More

Link

File # (7) ENTER

Next

File # (8) ENTER

Etc until all file numbers have been entered then:

ESC

Prefabricated Methods

Five prefabricated Methods, to run or to adapt, are present in the 402’s memory as file numbers 90 to 94. Although the tasks are the same, the parameters vary according to whether you have a Single- syringe or a Dual-syringe 402.

Configuration

Standard equipment with a 2 mL transfer tubing.

Left-hand (or only) syringe: 5 mL.

Right-hand syringe (if fitted): 500 µL.

Default values are used for parameters not listed.

 

 

Value for use with

File Number

TASK/Parameter

Dual-syringe Model

Single-syringe Model

Method 90

Dilute Normal

 

 

 

Dilution Ratio

10

10

 

Sample Volume

200

1000

 

Rinse

300

1500

 

 

 

 

Method 91

Dilute Serially

 

 

 

Total Volume

3000

3000

 

Sample Volume

100

1000

 

Nb of Dilutions

3

3

 

 

 

 

5-28

Page 78
Image 78