Image 117
Appendix D-3

Error Messages

Appendix D

Error Message

Cause/Possible Action

 

 

 

Parameters incompatible!

Refer to page 5-16.

 

 

 

Ratio impossible

Dilution ratio calculated by software is not valid; check

 

parameters using formulae given in page 5-18 (DILUTE task)

 

and correct one or both of the other values.

 

 

 

Ratio missing

Re-enter Method, supplying the missing parameter, when

 

prompted to do so (see DILUTE task).

 

 

 

Reservoir not allowed

Liquids must be aspirated using the probe.

 

 

 

Right syr life exhausted!

For the Dual syringe model: piston or right-hand syringe may

 

need servicing (refer to page 6-6).

 

 

 

Sample too small

Enter a larger value for Sample Volume. Verify parameters in

 

the task.

 

 

 

Sample volume missing

Re-enter Method, supplying the missing parameter, when

 

prompted to do so (see DILUTE task).

 

 

 

Syr capacity too small

Check volumes used in the task correspond to the configured

 

syringe size(s).

 

 

 

Syringe Life Exhausted!

For the Single syringe model: piston syringe may need

 

servicing (refer to page 6-6).

 

 

Total volume is missing!

Re-enter Method, supplying the missing parameter, when

 

prompted to do so (see DILUTE task).

 

 

Total volume missing

Re-enter Method, supplying the missing parameter, when

 

prompted to do so (see DILUTE task).

 

 

Total volume too small

Enter a larger value for the Total Volume, or check the volumes

 

used in the task.

 

 

Valve life exhausted!

Valve may need cleaning or servicing (refer to page 6-2).

xxxexists, overwrite?Select another file number, unless you want to overwrite file xxx.

Error Messages

Appendix D-3

Page 117
Image 117
Gilson 402 manual Error Messages, Appendix D-3