Gilson 402 manual Installation Removing the Cover, Connecting the Flat Cable

Models: 402

1 124
Download 124 pages 13.02 Kb
Page 35
Image 35
Installation

Installation

Removing the Cover

1)Undo the four screws (with two washers) at the base of the 402 that secure the top cover. Place the 402 in the upright position, pull the sides of the cover outwards, and then lift off the cover.

2)Undo the two nuts inside the front panel (near the top) that secure the blanking plate, then remove the blanking plate.

Connecting the Flat Cable

There is a flexible flat cable inside the 402 (attached to the lower chassis with Velcro) that connects the logic board to the Valve module. Detach the cable from the lower chassis, then use the free connector to connect to the motor drive board on the upgrade module. With reference to the opposite figure, the loose-end of the flat cable (with the second connec- tor) and the connecting cable may be re-attached, using their Velcro strips.

1.Hold the module upright, in front of the 402 front panel.

2.Thread the flat cable through the aperture for the module and connect it to the syringe motor drive board on the Tee module.

3

Upgrading to Dual-syringe Model

Second connector is not used

To tee module

3-11

Page 35
Image 35
Gilson 402 manual Installation Removing the Cover, Connecting the Flat Cable, Upgrading to Dual-syringe Model