Upgrading to Dual-syringe Model

Installation

Upgrading to Dual-syringe Model

3

Upgrading to Dual-syringe Model

The single syringe 402 can be upgraded to improve the versatility of the system.

To upgrade to the Dual-syringe 402, you must install a Tee-junction module (kit reference F4015010) at the right-hand-side of the 402. You must first install the syringe module (includes motor etc.) and then fit the Tee-junction.

Before installing the module: make sure that the ON/ OFF switch is in the OFF position, remove the power cord, and disconnect any hydraulic connections from the 402.