Gilson 402 Syringes, Flow Rates, Configuration Screen, Operation - Standard Use, Essential Steps

Models: 402

1 124
Download 124 pages 13.02 Kb
Page 41
Image 41
Syringe(s)

Operation - Standard Use

If you want to change the configured volume, press Chge (until you reach the correct value); when the volumes are the same press ENTER, press the ESC key to return to the Configuration Screen, or press Next, after which you see the Air Gap Volume Screen.

Key in the ‘Air gap volume’, press ENTER to confirm then press the ESC key to return to the Configuration Screen (see Chapter 5, for a definition of Air Gap).

Syringe(s)

You must check that the configured volumes are the

Syringe Screen

same as the sizes to be fitted. Press the Syr soft-key.

 

The use of the soft-keys is as follows:

 

Next is used to go to the next screen,

 

Prev is used to go to the previous screen,

 

Corr is used to supply a correction factor (see

 

Chapter 5),

 

Chge is used to select the size(s) of syringe(s),

 

Exch is used to lower the piston rods.

 

To change a syringe size, press the Chge soft-key

 

until you reach the corresponding volume, then

 

press ENTER to confirm. Press Exch if you are

 

ready to lower the piston rods, or when you have

 

two syringes, press Next. Check that the configured

 

volume for the right syringe is the same as the size

 

to be fitted. Press the Chge soft-key until you reach

 

the corresponding volume, then press ENTER to

 

confirm. Press Exch if you are ready to lower the

 

piston rods or press the ESC key to return to the

 

Configuration Screen.

 

Whenever you change the configured volume, the

 

software reminds you to ‘Exchange Syringes’ before

 

allowing you to quit the configuration screen. Go

 

directly to page 4-7, which explains how the ‘Ex-

 

change Syringes’ function works, unless you need

 

to change flow rates, etc.

 

Flow Rates

Flow rates should be specified according to the

 

sizes of syringe and tubing fitted to the 402. Refer

 

to Chapter 5 for more details, the maximum values

 

are given in the Tables, these are used as default

Air Flow Screen

values by the software.

 

Press the Flow soft-key; the Air Flow Screen appears.

 

4

Essential Steps

4-5

Page 41
Image 41