Appendix A: Perfo rmance Specification
A-17
1760 SeriesComponent Waveform/Vector Monitor User Manu al
TableA -19: Certifications and compliances (cont.)
CATEGORY STANDARDSOR DES CRIPTION
AdditionalCompliance IEC61010-1
Safety requirementsfor electrical equipment for
measurement,control, and laboratory use.
Safety Certification C ompliance
Temperature,
operating
+5 to +40_C
Altitude(maximum
operating)
2000 meters
EquipmentType Testand measuring
Safety Class Class1 (as defined in IEC 1010-1, Annex H) --grounded product
OvervoltageC ategory OvervoltageCategory II (as defined in IEC 1010-1, Annex J)
PollutionDegree PollutionDegree 2 (as defined in IEC 1010-1). Note: Rated for
indoor useonly.