Appendix B: Rem ote Connectors
1760 SeriesComponent Waveform/Vector Monitor User Manu al
B-4
Remote ConnectorConverterIf the 1760--Seriesis used as a direct replacement for a
1740--Series or 1750--Series instrument, construct an adapter
or rewire the remote cable according to Figure B-2. If the
1760--Series replaces an Option 16 i nstrument, construct the
adapter in the same manner,omitting the pin 3-to-pin-25
connection.
13
25
RGB INPUT
RGB ENABLE*
REMOTESYNC IN
GROUND
REMOTESYNC ENABLE
3
19
6
7
5
Remote Plug
14
1
1760--Series
Rear-Panel
REMOTE Connector
25
8
9
10
1
14
13
* (Do notconnect when replacing Option 16 instruments.)
Figure B-2. Usingthe 1760 ---Series as a 1740/1750---Series replacement.